Yes. You will have to Rename the DAQ Hardware as described below.


Modal Amplified Software – Renaming Hardware

 

1) Install Modal Amplified software - it is recommended that you do not attach modal hardware/cabling to the PC during installation

 

2) Connect hardware to PC - Launch NI Maxx

 

After installation of Modal Amplified, connect modal hardware to PC.  When encountering RDI modal hardware for the first time, you will see a National Instruments menu appear as shown below.  Select “Go” to configure the hardware.  If this message is not seen, manually launch NI Max software using Windows search bar.

 

[Display properties might need to be adjusted to see the small font!]

 

Graphical user interface, application

Description automatically generated

 


 

 

Note!  If NI Max software is very small, try the following:

  • Increase scaling factor of your display temporarily or…
  • Close NI Max, then:
    1. Using File Explorer, navigate to C:\Program Files (x86)\National Instruments\MAX\
    2. Right-click on NIMax.exe, select Properties->Compatibility->Change high DPI settings.  
    3. Select Override high DPI scaling behavior/Scaling performed by:  
    4. Choose System from drop down.

Graphical user interface, text, application

Description automatically generated

 


 

 

Expand the Tree in upper left corner – again, expand “Devices and Interfaces” as shown.

 

Graphical user interface, text, application

Description automatically generated

 

 

Note – if you see multiple NI cDAQ-9174 entries, simply right-click/delete entries with the red ‘X’, as shown.

 

Graphical user interface, text, application

Description automatically generated

 

 


 

 

Renaming the active NI hardware:  if needed, rename the DAQ itself as shown below.  Enter into “Name” field cDAQ1…then select Save.  If already correctly named, skip this step! 

 

Graphical user interface, application

Description automatically generated

 

 

Drop down to the 9230 Card; rename as shown – enter AI1 [Analog In 1], then select Save.

 

Graphical user interface, application

Description automatically generated

 

Drop to the 9402 Card; rename as shown – enter DIO1 [Digital In/Out 1], then select Save.

 

Graphical user interface

Description automatically generated

 

Shaker card:  if the Shaker card is present, continue the renaming process for this module; rename that card AO1 (Analog Out 1), then select Save.

 

Upon completion, NI Max should reflect the following (ignore third card, 9260 if no Shaker card was purchased)

 

 

Graphical user interface, text, application, chat or text message

Description automatically generated

 

 

 

 

Exit NI Max – Modal Amplified is ready to go!  

As long as your hardware doesn’t change, this process will not be required again.

 
 

 

Troubleshooting:

 

  • If hardware reflected within NI Max does not appear as described above, power down hardware and remove Modal-to-USB from PC
    1. Relaunch NI Max
  • If hardware cannot be found, check Device Manager as shown below:  attempt to update Driver

 

Graphical user interface, application, Word

Description automatically generated


 

  • If a rogue entry exists, such as shown below, delete it. 

Graphical user interface, text, application

Description automatically generated

 

Table

Description automatically generated




Related Articles:


What kind of Modal Hammer can I purchase to work with Modal Amplified™?

Do I need to do anything special to use my New Feature (Iris MX™, High Speed Plugin, Iris M™ Tachometer, Iris M Color Camera, Modal Amplified™, and Iris Explorer™ (Spot® Payload)) for the first time?

My Iris MX™ download is slow or generates error messages. What can I do to fix this?